Monitor Installation and Patient Safety 11-5

M

onitor

Ins

ta

lla

ti

on

a

nd Pa

ti

e

n

t Sa

fe

ty

Installation Information

Power
Source
Require-
ments

The CMS systems and the Anesthetic Gas Module can be operated from
an AC source of 100 - 120V

±

10%, or 220 - 240V

±

10%,

50 - 60Hz.

Typical input currents for the monochrome and color systems are:

Typical input current for the Anesthetic Gas Module is:

100 - 240V = 1.0 amps

Typical input current for the V24 and V26 is:

90 - 250V = 150VA

Grounding
the System

To protect the patient and hospital personnel, the cabinet of the installed
equipment has to be grounded. The equipment is supplied with a
detachable 3-wire cable which grounds the instrument to the power line
ground (protective earth) when plugged into an appropriate 3-wire
receptacle. If a 3-wire receptacle is not available, consult the hospital
electrician.

Warning

Warning

Do not use a 3-wire to 2-wire adapter.

Monochrome CRT

100 - 120V = 2.2 amps
200 - 240V = 1.2 amps

Color CRT

100 - 120V = 2.6 amps
200 - 240V = 1.4 amps

Color Flatscreen

110 - 120V = 2.0 amps
200 - 240V = 1.0 amps