Data Transfer Module

Data Transfer 10-5

D

a

ta

T

ra

n

sf

er

When the module is unplugged, it retains data for at least one hour.

After data is transferred to the monitor, the module is automatically
erased. This is to ensure that the module is empty when it is used for the
next transfer.

Symbols to
Indicate Key
Functions

As detailed in the table below, some versions of this module now use
symbols, instead of words, to indicate the function of some keys. If the
monitor’s Reference Manual tells you, for example, to select the Monitor
key, you should press the key marked with the corresponding “monitor”
symbol.

The design change also means that you will now find the module’s
product number on the rear of the module’s housing, not on the front.

What is
Transferred

The module is capable of storing and transferring up to 24 hours of trend
data and demographic information. When a transfer is initiated, the
following data is transferred:

Continuously monitored parameters (for example, ECG)

Resampled Values

Aperiodic measurements (for example, CO)

All patient demographic information except bed number

The time at which calculations were performed in the
Hemodynamic, Oxygenation, and Ventilation calculation task
windows

Marked events

ST events (lead changes, reference changes)

The following data is not transferred:

Waveforms, including ST and CO waveforms

Split Screen trends

Arrhythmia events