Using an ITE Display

The CMS and V24 and V26 Patient Monitors 1-39

EMC

1. The display must either fulfill the requirements of the EN60601-1-2

{IEC 601-1-2}, or

it must conform to the requirements of the EMC standard for ITE
devices, EN50082-1: 1997 and EN50081-1 / CISPR 22.

2. The video cable between M1167A/77A CMS Patient Monitoring

System and the ITE display must not exceed a length of 3.0m. For
cables exceeding 3.0m EMC testing according to IEC 801-4:1988 has
to be done.

Perfor-
mance
Require-
ments

Philips cannot assure compliance with the ANSI/AAMI EC-13 Standard for
Cardiac Monitors, Heart-rate Meters and Alarms when using ITE displays.
Compliance with the ECG aspect ration and 25 mm/s

±

10% sweep-speed

specifications can only be assured when using the Philips Displays
M1092A, M1094B and M1095A. In addition, the following display
specifications are recommended:

Specification

Requirement or Value

Units

Resolution (pixel addressability)

1024 x 768

Dots x lines

Vertical Refresh Rate

60

Hz (non-interlaced)

Red, Green and Blue Video Inputs

~0.7

V p-p

Vertical & Horizontal Multi-Sync Inputs

5

V TTL

Video-Cable Connector

HD15male