Parameter Settings Transfer Messages

Setting up your Monitor 3-63

S

e

tt

ing up y

our

M

onit

o

r

Parameter Settings Transfer Messages

The following messages related to Parameter Settings Transfer can
appear on the display:

Message

Condition

Action required

(Parameter name)
no settings
transfer capability

The indicated parameter
module is not a T
module.

Insert a module that has
a T on the front.

(Parameter name)
check settings''

No settings are stored in
the module, or the
settings are incorrectly
stored.

Enter the parameter
Task Window and check
the settings. The
settings are
automatically stored
when you exit the Task
Window.

Check Patient
Category and
(parameter name)
settings

The patient category
stored with the settings
in the module is
different from that of the
monitor.

Either

change the

patient category for the
monitor or change the
parameter settings.

(Parameter name)
unable to store
settings in module

The parameter module
was removed before the
settings could be stored,
or there is a malfunction
in the module. (This
message can appear with
Settings Transfer On or
Off.)

Press ,

re-insert the module and
check/adjust the
settings in the
parameter Task
Window. Exit the Task
Window and leave the
module in the rack for
about 10 seconds before
you remove it. If the
same message appears,
replace the module.

Silence Reset