background image

Loading Paper

Recording Functions 6-55

Note—

The first sheet will come out folded, as shown.

Cleaning the
Roller on
the Four
Channel
(M1117A)
Recorder

If the paper doesn’t move through the roller properly, a possible cause is
that the roller needs to be cleaned. (Be sure your hands are clean before
touching the roller).

Equipment

Required

Lint-free

cloths or wipes

Alcohol solvent that is low in water content. We recommend either
reagent grade denaturated ethanol or isopropanol (90% rather than
70%)

Directions

1. Turn the power off.

2. Open the door and remove any paper from the recorder.

3. Wipe the visible part of the roller with a saturated (but not dripping)

cloth. Rotate the roller with clean fingertips to expose the entire
surface and wipe the other side.

4. Remove any excess liquid by wiping with a dry cloth.

Caution

The roller must be completely dry before using the recorder or you can
permanently damage the recording mechanism.