Using Philips Patient Care System with an Arrhythmia Computer

Other Patients 4-13

Ot

h

e

r Pa

ti

e

n

ts

Arrhythmia Alarms on the 3150A/3153A Arrhythmia Computer

Message

Minimum Condition Requiring an Alarm

*** ASYSTOLE

No QRS for 4 consecutive seconds

*** V-FIB/TACH

Fibrillatory wave for 4 consecutive seconds

*** V-TACH

Consecutive PVCs >/= V-Tach Run limit and
HR > V-Tach HR limit

*** TACHY yyy > xxx

Heart Rate (yyy) greater than the Extreme
Tachy limit (xxx)

*** BRADY yyy < xxx

Heart Rate (yyy) less than the Extreme
Brady limit (xxx)

** NON-SUSTAIN VT

A run of Vs having a ventricular HR>V-Tach
HR limit, but lasting for less than the V-Tach
Run limit

** VENT RHYTHM

A dominant rhythm of adjacent Vs > vent
rhythm limit and ventricular HRlimit

** RUN PVCs

Run of PVCs greater than 2

** PAIR PVCs

Two consecutive PVCs between non-PVCs

** PACER NOT CAPT

No QRS for 1.75 x the average R-R interval
with Pace Pulse (paced patient only)

** PACER NOT PACE

No QRS and Pace Pulse for 1.75 x the
average R-R interval (paced patient only)

** PAUSE

No beat detected for 1.75 x average R-R
interval for HR <120, or no beat for 1 second
with HR >120

** SVT

Run of SVPBs >/= SVT Run limit and with
SVT Heart Rate greater than the SVT HR
limit