beautypg.com

Event review – Philips V24CT User Manual

Page 285

background image

Viewing Neonatal Events

Neonatal Event Review 9-11

Ne

ona

ta

l Ev

e

n

t

Revi

ew

Example of an Event Summary Line

The above example of an event summary line, conveys the following
information:

A total of 19 neonatal events are stored.

Four Apnea events occurred, of which two resulted in a
Desaturation.

Three isolated Bradycardia events occurred, none of them resulted
in a Desaturation.

One associated Apnea and Bradycardia occurred which also
resulted in a Desaturation.

Nine isolated Desaturations occurred.

Two events were manually captured and stored.

Event Storage

If the number of stored events reaches a total of 44

1

, the following status

message is displayed in the Event Review task window:

“Event Storage>90% full”

If the number of stored events exceeds a total of 48

2

, the oldest events are

overwritten and the following status message is displayed in the Event
Review task window:

“Event Storage overflow. First events overwritten”

Event Review

19 Stored Events: A(D): 4(2) B(D): 3(0) AB(D): 1(1) D: 9 M: 2

1. 22 for V24 and V26 Patient Monitor

s

2. 24 for V24 and V26 Patient Monitor

s

This manual is related to the following products: