beautypg.com

General disinfecting of the system, General disinfecting of the system -4 – Philips V24CT User Manual

Page 370

background image

General Disinfecting of the System

13-4 Maintenance

M

a

int

e

na

nc

e

General Disinfecting of the System

We recommend that you disinfect the product only when necessary as
determined by your hospital’s policy, to avoid long term damage to the
product.
We also recommend that the products being disinfected be cleaned first,
as described under “General cleaning of the System” on page 13-2.
Use the recommended disinfecting substances listed below:.

Caution

To avoid damage to the product, observe the following general
precautions for disinfection. You should only deviate when this is
explicitly stated in the disinfecting instruction of a specific product.

1. Do NOT use Povodine

®

, Sagrotan

®

, Mucocit

®

disinfecting agents

or strong solvents e.g. acetone.

2. Always dilute according to the manufacturer’s instructions or use

lowest possible concentration.

3. Do not allow any liquid to enter the case.

4. Never submerge any part of the system.

Alcohol
based

Ethanol 70%, Isopropanol 70%

Aldehyde
based

Cidex

Bleach

Dilution of sodium hypochlorite (laundry bleach): concentration
ranging from 500ppm (1:100 dilution of household bleach) to
5000ppm (1:10 dilution of household bleach),
Hydrogen peroxide 3%, Chlorox

(1:10 dilution),

Dakin’s Solution.

Phenol based

Wofasept

, Sporicidin

.