beautypg.com

Operating instructions, Operating instructions -3 – Philips V24CT User Manual

Page 357

background image

AC and DC (Battery) Operation

Battery Information (V24CT and V26CT only) 12-3

Ba

tte

ry

I

n

for

m

a

ti

on

(V

24

CT

an

d

V

26C

T

Operating
Instructions

The following is a list of AC and battery operating instructions.

The batteries continuously charge when the instrument is
connected to AC power. Charging occurs whether the monitor is
turned Off or On.

Caution

Philips Medical Systems recommends you fully charge the batteries
before using the V24CT for the first time. Plug the monitor into AC power
(line power) for an initial charging cycle.

AC power is indicated by the

green LED indicator.

Battery charging is indicated by the

amber LED.

The LED indicator flashes when the batteries first start charging,
and continues flashing until the batteries are 30 - 40% charged.

When the batteries are 30 - 40% charged, the indicator goes from
flashing to continuously on. The

indicator will be

on steady until the batteries are 90% charged. The

indicator will turn off when both batteries are 90% charged or when
the line power is disconnected.

The

green LED illuminates when the battery

charging function has reached 90% charged, and will stay on steady
even when the batteries are fully charged. If the line power is
disconnected, this LED indicator will be turned off.

AC Power

Battery Charging

Battery Charging

Battery Charging

Battery Charged

Delayed
Record

Trends/Calc

Module Setup

Confirm

Battery
Charging

Battery
Charged

AC
Power