Page 327

Installation Information
Monitor Installation and Patient Safety 11-11
M
onitor
Ins
ta
lla
ti
on
a
nd Pa
ti
e
n
t Sa
fe
ty
Philips M1205A V24CT and V26CT
Condensa-
tion
Make sure that during operation, the instruments are free of
condensation. Condensation can form when equipment is moved from
one building to another, thus being exposed to moisture and differences
in temperature.
Warning
Warning
Possible explosion hazard if used in the presence of flammable
anesthetics.
Warning
Warning
Do not use cellular phones in the vicinity of the system. Cellular
phones may generate excessive radiated fields which can disturb
the specified function of the system.
Operating
Storage
Temperature Range
5 to 35
°
C (41 to 95
°
F)
-15 to 40ºC (5 to 104ºF)
Humidity
up to 95% RH at 35
°
C (95
°
F)
up to 90% RH at 40ºC (104ºF)
Altitude Range
up to 4,600m (15,000 ft)
up to 4,600m (15,000 ft)