Troubleshooting

Data Transfer 10-19

D

a

ta

T

ra

n

sf

er

Troubleshooting

Caution

In a centrally networked system, if the previous patient is discharged only
at the monitor and a new patient's data is transferred to the monitor, the
new patient's name and ID will be overwritten by the central station's
broadcast of the previous patient's name. To prevent this situation from
occurring, the previous patient should be discharged from central station
before

the transfer.

The following table outlines the Data Transfer behavior under various
events.

Event

Results

Power Off < 1 min

Transfer will restart automatically

Power Off > 1 min

Transfer is stopped. Collection of new data
must be restarted manually.

Op. Mode Change

Transfer is stopped. Data in module remains
intact.

Config Mode Change

Transfer is stopped. Data in module remains
intact.

Patient Discharge or End
Case

Transfer is stopped. Data in monitor is
deleted. Data in module remains intact.