beautypg.com

Operating controls, Operating controls -16 – Philips V24CT User Manual

Page 290

background image

Viewing oxyCRG Episodes

9-16 Neonatal Event Review

Ne

ona

ta

l Ev

e

n

t

Revi

ew

The exact time an event was triggered is indicated by a vertical line
which is positioned on the graph according to the monitor's
configuration. Depending upon the user's requirements, the monitor can
be configured (in Configuration Mode) to display and store an oxyCRG
Episode in the following way:

Two minutes of Pre-Event Time / Two minutes of Post-Event Time
(the default) or

One minute of Pre-Event Time / Three minutes of Post-Event Time

Note—

In both cases manual events are displayed with four minutes Pre-

Event Time.

The association of events is determined by the configuration of the post
event time. The events that occur during the configured two or three
minutes of post event time are associated and displayed as one event in
the Event Review and oxyCRG Episode Task Window.

Operating
Controls

In the oxyCRG task window you can:

1. Select the previous episode.

2. Select the next episode.

3. Return to the Event Review task window.

4. Record an episode.

5. Print an episode.

6. Delete an event.

Selecting an

Episode

Selecting the

and/or

softkeys

jumps from one oxyCRG episode to the next in both directions.

Entering the

Event Review

Task Window

Selecting the

softkey returns you to the Event Review

task window. The event last selected in the oxyCRG Episode task
window will be selected here.

Previous Episode

Next Episode

Event Review