Alarm Display

Alarm Functions 5-3

Ala

rm Func

tions

Alarm Lamps

Red and yellow alarm lamps on the control panel signify the severity of
the cause of the alarm. One of these lamps is always lit in the event of an
alarm sound. In the case of more than one alarm, the highest severity
alarm lamp is always lit.

What the

Symbol Means

When the alarm capability is suspended by using the

softkey, or the

hardkey, the

sign on the control panel is

illuminated. If, however, you have turned the individual parameter alarms
off one by one, the

sign on the control panel is not illuminated, but a

appears next to the individual parameter numerics on the screen.

Suspending
Alarms

There are four alarms-off modes available. You configure which one you
want at installation of your system. The four modes are:

1. ALARMS SUSPND 1 MIN
2. ALARMS SUSPND 2 MIN
3. ALARMS SUSPND 3 MIN
4. ALARMS SUSPENDED

ALARMS SUSPENDED 1, 2, 3 MIN

In these modes the alarm capability can be switched off, but is
automatically switched on after 1, 2 or 3 minutes (unless you have
switched it on manually beforehand with the

softkey

in the Alarms Selection Window, or the

hardkey).

ALARMS SUSPENDED

In this mode the alarm capability is switched off permanently and must be
switched on manually when required (using the

softkey in the Alarms Selection Window, or the

hardkey).

Note—

For safety reasons, we recommend that you always configure your

monitor to ALARMS SUSPND 1, 2 or 3 MIN for Intensive Care as well as
Operating Room use. If you want to avoid nuisance alarms caused by
artifacts in the Operating Room, configure your alarm limits carefully or
switch off individual paramter alarms, applying special vigilance to these
parameters. All monitors within a unit should have the same alarm
configuration.

Suspend Alarms

Suspend

SwitchOn Alarms

Suspend

SwitchOn Alarms

Suspend