background image

Alarm Display

5-2 Alarm Functions

Ala

rm Func

tions

Alarm Display

The alarm messages are displayed across the top of the screen. When
you switch the power on, the system alarm capability is switched off.
This is indicated by an ALARMS SUSPENDED or ALARMS SUSPND 1, 2
or 3 MIN message at the top right corner of the screen, depending on the
alarm suspended mode which is configured on the monitor. The Alarm
Suspend symbol(

) on the control panel is also illuminated.

Note—

When you switch the monitor on, a loudspeaker test is conducted

for approximately 3 seconds. If there is no sound, please check the
volume settings.

If there is more than one red alarm, yellow alarm or INOP message, you
will see an up arrow (

) next to the message. The messages are

displayed in rotation every 2 seconds. If your patient has more than one
alarm, you can see a list of all these active alarms in the Alarm Messages
Task Window. To access this Task Window press

or

followed by

.

Warning

Warning

Alarms and INOPs are not active in the configuration and service
modes, but are active in monitoring and demo modes.

Alarm
Functions
on the
Control
Panel

Alarms/Volume

Alarms

Alarm Messages

Alarms/

Other

Monitor

Setup

Main

Screen

Change

Screen

Silence/

Reset

Alarm Lamps

Alarms Suspended Lamp

Suspend

Silence/Reset key

Suspend key

Volume

Patients