beautypg.com

Heart rate (hr) / pulse, Resp, Pressure – Philips V24CT User Manual

Page 125: Spo2

background image

Changing the Patient Category

Setting up your Monitor 3-51

S

e

tt

ing up y

our

M

onit

o

r

Heart Rate
(HR) / Pulse

The expected range can be estimated by performing a manual pulse
count.

RESP

In general, use the Adult or Pediatric category if rates below 50 rpm are
expected; use Neonatal if expected rate is above 100.

Pressure

The algorithm responsible for respiration artifact suppression is used in
all three patient categories for the 10 and 30 scales. In the Pediatric and
Adult categories, the algorithm is also used for the 60 scale.

SpO

2

Default for averaging time is 10 seconds for Adult, Pedi and Neo.

Note—

Default can be changed in Configuration Mode.

Adult (bpm)

Pedi/Neo (bpm)

Cardiotach Range

15 to 300

15 to 350

HR Alarm Limit Range

15 to 250

15 to 300

Pleth/Pressure Pulse Alarm Limit
Range

30 to 250

30 to 300

Adult/Pedi (rpm)

Neo (rpm)

Respirotach Range

0 to 120

0 to 180

High Alarm Limit Range 10 to 100

30 to 150

Low Alarm Limit Range

0 to 95

0 to 145

This manual is related to the following products: