Heart rate (hr) / pulse, Resp, Pressure – Philips V24CT User Manual
Page 125: Spo2

Changing the Patient Category
Setting up your Monitor 3-51
S
e
tt
ing up y
our
M
onit
o
r
Heart Rate
(HR) / Pulse
The expected range can be estimated by performing a manual pulse
count.
RESP
In general, use the Adult or Pediatric category if rates below 50 rpm are
expected; use Neonatal if expected rate is above 100.
Pressure
The algorithm responsible for respiration artifact suppression is used in
all three patient categories for the 10 and 30 scales. In the Pediatric and
Adult categories, the algorithm is also used for the 60 scale.
SpO
2
Default for averaging time is 10 seconds for Adult, Pedi and Neo.
Note—
Default can be changed in Configuration Mode.
Adult (bpm)
Pedi/Neo (bpm)
Cardiotach Range
15 to 300
15 to 350
HR Alarm Limit Range
15 to 250
15 to 300
Pleth/Pressure Pulse Alarm Limit
Range
30 to 250
30 to 300
Adult/Pedi (rpm)
Neo (rpm)
Respirotach Range
0 to 120
0 to 180
High Alarm Limit Range 10 to 100
30 to 150
Low Alarm Limit Range
0 to 95
0 to 145