Changing the Patient Category

Setting up your Monitor 3-49

S

e

tt

ing up y

our

M

onit

o

r

1

The figure given for overpressure protection is the maximum pressure

the cuff can reach for 2 seconds before it deflates.

NBP Recommendations

Philips Medical Systems makes the following recommendations when
selecting a patient category for NBP:

1. Choose the category with the most appropriate safety values for

your patient.

2. Reduce the stress on the patient by choosing the appropriate cuff

size (a cuff that is too large prolongs inflation time).

3. Use a repetition time appropriate to the patient.

ECG

The following factors can cause the cardiotach to calculate an incorrect
heart rate:

Morphology of the ECG,

Weak signal,

Paced patients.

Morphology of

the ECG

Philips offers two different ECG algorithms to allow for differing
morphologies.

Max. deflation time

80s

80s

60s

Max. measurement time

100s

100s

60s

Overpressure protection

1

300 mmHg

300 mmHg

150 mmHg

Cuff Size

A to E

Any Cuff

1 to 4 & A

Default Inflation Pressure

165 mmHg

125 mmHg

100 mmHg

NBP Safety Values for Differing Patient Categories

Safety Value

Patient Category

Adult

Pedi

Neo