Changing the Patient Category
Setting up your Monitor 3-49
S
e
tt
ing up y
our
M
onit
o
r
1
The figure given for overpressure protection is the maximum pressure
the cuff can reach for 2 seconds before it deflates.
NBP Recommendations
Philips Medical Systems makes the following recommendations when
selecting a patient category for NBP:
1. Choose the category with the most appropriate safety values for
your patient.
2. Reduce the stress on the patient by choosing the appropriate cuff
size (a cuff that is too large prolongs inflation time).
3. Use a repetition time appropriate to the patient.
ECG
The following factors can cause the cardiotach to calculate an incorrect
heart rate:
•
Morphology of the ECG,
•
Weak signal,
•
Paced patients.
Morphology of
the ECG
Philips offers two different ECG algorithms to allow for differing
morphologies.
Max. deflation time
80s
80s
60s
Max. measurement time
100s
100s
60s
Overpressure protection
1
300 mmHg
300 mmHg
150 mmHg
Cuff Size
A to E
Any Cuff
1 to 4 & A
Default Inflation Pressure
165 mmHg
125 mmHg
100 mmHg
NBP Safety Values for Differing Patient Categories
Safety Value
Patient Category
Adult
Pedi
Neo