Full calibration – Super Systems 9205 Series User Manual
Page 121
![background image](https://www.manualsdir.com/files/806366/content/doc121.png)
Series 9205 Operations Manual Rev A
120
box. If the input was showing 1748°F, while sourcing 1750°F, then the “Offset” difference would be 2°F to
bring the display back up to 1750°F.
Note: The current cold junction value is for display purposes. It will need to be used when determining the
cold junction offset. You will need to add to or subtract from the displayed value the amount of difference,
example 2°F difference being applied to 90 (like in the image) your new CJ value will need to be 92°F.
When ready click on the Calibrate button to begin the calibration.
Wait 120 seconds and verify with a source calibration device with the correct T/C type. Note: During a
normal calibration procedure, the user should zero and span calibrate all of the inputs first and then
perform a cold junction calibration, if necessary.
Click on the Done button when finished. You will then be able to verify on the main display your sourced
value when compared to the actual reading in Configurator.
Full Calibration
NOTE: Full calibration is performed by SSi. The below information is provided for reference only.
The
Calibration
screen for the Full Calibration menu option is identical in function and layout to the User
Calibration’s
Calibration
screen. The only difference is the Full Calibration’s
Calibration
screen contains
more screens. This list of screens is listed below in sequential order. Screens only found in the Full
Calibration menu option are displayed in
Italics
. Screens only found in the User Calibration menu option
are displayed in bold.
1.
Cold Junction
2.
Zero/Span Input 0/Range 0
3.
Zero/Span Input 0/Range 1
4.
Zero/Span Input 0/Range 2
5.
Zero/Span Input 0/Range 3
6.
Zero/Span Input 1/Range 0
7.
Zero/Span Input 1/Range 1
8.
Zero/Span Input 1/Range 2
9.
Zero/Span Input 1/Range 3
10.
Zero/Span Input 2/Range 0
11.
Zero/Span Input 2/Range 1
12.
Zero/Span Input 2/Range 2
13.
Zero/Span Input 2/Range 3
14.
Zero/Span Input 3/Range 2
15.
Zero/Span Input 1 Range Jumper
16.
Zero/Span Input 2 Range Jumper
17.
Zero/Span Input 3 Range Jumper
18.
Zero/Span Output 1
19.
Zero/Span Output 2