beautypg.com

Super Systems AC20 User Manual

Page 80

background image

78

R.H05 -

R.H05 -

R.H05 -

R.H05 -

R.H05 -

Time remaining to the next automatic

Time remaining to the next automatic

Time remaining to the next automatic

Time remaining to the next automatic

Time remaining to the next automatic

Burnoff

Burnoff

Burnoff

Burnoff

Burnoff
This is a reading parameter only and it is available when "bF.tr" is
different from "no.tr".
Upper display: Gr.8
Lower display: bF.ñn
Range: from 1.00 to 24.00 hh. mm.

When a primary variable different from “O2” is selected.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: if a power down occurs, this information will be lost,

when power is restored, the countdown will restart from
the "bF.tr" [R.H03] value.

R.H06 -

R.H06 -

R.H06 -

R.H06 -

R.H06 -

Time remaining to the next automatic

Time remaining to the next automatic

Time remaining to the next automatic

Time remaining to the next automatic

Time remaining to the next automatic

Probe test

Probe test

Probe test

Probe test

Probe test
This is a reading parameter only and it is available when "pb.tr" is
different from "no.tr".
Upper display: Gr.8
Lower display: Pb.ñn
Range: from 1.00 to 24.00 hh. mm.

When a primary variable different from “O2” is selected.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: if a power down occurs, this information will be lost,

when power is restored, the countdown will restart from
the "Pb.tr" [R.H04] value.

R.H03 -

R.H03 -

R.H03 -

R.H03 -

R.H03 -

Automatic Burnoff interval

Automatic Burnoff interval

Automatic Burnoff interval

Automatic Burnoff interval

Automatic Burnoff interval

Upper display: Gr.8
Lower display: bF.tr
Range: from 1.00 to 24.00 hh. mm.

Above this 24.00, the display shows "no.tr" (no time
interval is programmed)

When a primary variable different from “O2” is selected.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: the time setting may be modified any time but it will be

performed at the next Burn off, exception made for the
changement from/to "no.tr" that it will be immediately
activated.

R.H04 -

R.H04 -

R.H04 -

R.H04 -

R.H04 -

Automatic probe test interval

Automatic probe test interval

Automatic probe test interval

Automatic probe test interval

Automatic probe test interval

Upper display: Gr.8
Lower display: Pb.tr
Range: from 1.00 to 24.00 hh. mm.

Above this 24.00, the display shows "no.tr" (no time
interval is programmed)

When a primary variable different from “O2” is selected.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: the time setting may be modified any time but it will be

performed at the next Probe test, exception made for the
change from/to "no.tr" that it will be immediately
activated.