Caution – Measurement Computing LogBook Series User Manual
Page 31

QS300
-6
LogBook/300 Quick Start Guide
Device Name is displayed with the default name, numbered successively as configured. This field can
be changed to any descriptive name as desired.
Connection Type can be serial or parallel port.
Device Connection specifies the port name.
Protocol is used to set the parallel port protocol (ECP only) or serial protocol (RS-232 or RS-485).
Device Timeout specifies the number of seconds LogView will be wait for a LogBook response before
displaying an error condition.
ECP Parallel Port
To use parallel port communication with an attached LogBook/300, your PC must
support the ECP protocol AND be set in the ECP mode.
The majority of today’s PCs support the Enhanced Computer Port protocol (ECP). If your computer does
not support ECP, you can communicate with the LogBook/300 via the RS-232 serial port, or you can add an
ECP-compatible ISA board or PC-Card parallel port. Setting the PC to ECP mode varies with different
computers. On some computers, you can enter the BIOS Setup utility from Windows Settings or during
startup by pressing the F1 function key. The Parallel Port Mode property can be found under the Peripheral
Configuration group menu item. If necessary, consult your PC’s documentation or your PC’s manufacturer.
To ensure ECP compatibility after proper setup, use the Test Hardware utility
(described on page
QS300
-7). Before testing, make sure LogBook/300 is properly
connected, powered on, and that the Parallel Port Mode is set to ECP (in BIOS Setup).
CAUTION
Making errors in BIOS Setup can disrupt your system’s operation. If test hardware
indicates a problem and you have inadequate experience with the BIOS Setup utility,
consult your System Administrator or other qualified individual.