Measurement Computing LogBook Series User Manual
Page 159
![background image](https://www.manualsdir.com/files/797701/content/doc159.png)
Software Reference
989592
LogView LV-33
Analog Input Channel Configuration, 2-Point Cal Tab Selected
Calibration. The above figure shows the columns in the 2-Point Calibration tab. To fine tune the
accuracy of the value coming from a linear sensor, LogView provides both 1- and 2-point calibration.
One-point calibration can be used to zero a channel—as in a thermocouple, which is usually more
accurate in scale than offset. You might place the thermocouple in an ice bath and set just the 0°C point.
2-point calibration determines the scale and offset factors to convert the raw readings into accurate
calibrated readings. 2 points, of known (set) value, must be compared with 2 sensor (actual) readings.
For example, to calibrate a strain-gage scale, unload the scale, type 0.0 into the Set P1 column, then click
the Get P1
scale; type 100 into the Set P2 column; then click the Get P2
value (95.0). LogView automatically computes the Cal Offset factor (near 0) and Cal Scale factor (near
1). From now on, LogView automatically applies the y = mx + b calculation to the incoming reading to
produce the calibrated reading.