Cleaning fixed connectors, Mportant – EXFO PM-1600 High-Speed Power Meter User Manual
Page 98
Maintenance
88
PM-1600
Cleaning Fixed Connectors
Cleaning Fixed Connectors
Regular cleaning of connectors will help maintain optimum performance.
There is no need to disassemble the module.
To clean fixed connectors,
1. Fold a lint-free wiping cloth in four to form a square.
2. Moisten the center of the lint-free wiping cloth with only one drop of
isopropyl alcohol.
3. Gently wipe the connector threads three times with the folded and
moistened section of the wiping cloth.
4. With a dry lint-free wiping cloth, gently wipe the same surfaces three
times with a rotating movement.
5. Throw out the wiping cloths after one use.
I
MPORTANT
To help keep the various connectors and adapters clean, EXFO
recommends that you install protective caps when the unit is not in
use. You should also clean the fiber ends before every connection.
I
MPORTANT
Alcohol may leave traces if used abundantly. Avoid contact between
the tip of the bottle and the wiping cloth, and do not use bottles
that distribute too much alcohol at a time.
I
MPORTANT
Isopropyl alcohol takes approximately ten seconds to evaporate.
Since isopropyl alcohol is not absolutely pure, evaporation will
leave microscopic residue. Make sure you dry the surfaces before
evaporation occurs.
Maintenance
88
PM-1600
Cleaning Fixed Connectors
Cleaning Fixed Connectors
Regular cleaning of connectors will help maintain optimum performance.
There is no need to disassemble the module.
To clean fixed connectors,
1. Fold a lint-free wiping cloth in four to form a square.
2. Moisten the center of the lint-free wiping cloth with only one drop of
isopropyl alcohol.
3. Gently wipe the connector threads three times with the folded and
moistened section of the wiping cloth.
4. With a dry lint-free wiping cloth, gently wipe the same surfaces three
times with a rotating movement.
5. Throw out the wiping cloths after one use.
I
MPORTANT
To help keep the various connectors and adapters clean, EXFO
recommends that you install protective caps when the unit is not in
use. You should also clean the fiber ends before every connection.
I
MPORTANT
Alcohol may leave traces if used abundantly. Avoid contact between
the tip of the bottle and the wiping cloth, and do not use bottles
that distribute too much alcohol at a time.
I
MPORTANT
Isopropyl alcohol takes approximately ten seconds to evaporate.
Since isopropyl alcohol is not absolutely pure, evaporation will
leave microscopic residue. Make sure you dry the surfaces before
evaporation occurs.