beautypg.com

Sense:power:range:low, Sense:power:reference – EXFO PM-1600 High-Speed Power Meter User Manual

Page 145

background image

Remote Control (SCPI) Commands

High-Speed Power Meter

135

Specific Commands

SENSe:POWer:RANGe:LOW

Description This command sets the power measurement range to manual

low (for the currently selected channel when using a PM-1620).

If this command is sent while an acquisition is in progress, the

“Acquisition already running” error message will be raised.

Syntax

SENS:POW:RANG:LOW

Parameters The parameter can be 1, 2, 3, 4, 5, or 6

corresponding to manual LR1 to LR6.

Example

SENS:POW:RANG:LOW 4

See also

SENS:POW:RANG?, SENS:POW:RANG:AUTO,

SENS:POW:RANG:AUTO?, SENS:POW:RANG:HIGH, and

SENS:POW:RANG:LIST?

SENSe:POWer:REFerence

Description This command sets a user-defined reference value (in dBm or

W), for the selected channel when using a PM-1620. If this

command is sent while an acquisition is in progress, the

“Acquisition already running” error message will be raised.

Syntax

SENS:POW:REF

Parameters The parameter is the user-defined reference

value, in the “±99.999” (dBm) or “+999.9999E±99” (W) format.
The parameter represents the unit:
“DBM”sets dBm
“W”sets W

Example

SENS:POW:REF 11.23 dBm

Remote Control (SCPI) Commands

High-Speed Power Meter

135

Specific Commands

SENSe:POWer:RANGe:LOW

Description This command sets the power measurement range to manual

low (for the currently selected channel when using a PM-1620).

If this command is sent while an acquisition is in progress, the

“Acquisition already running” error message will be raised.

Syntax

SENS:POW:RANG:LOW

Parameters The parameter can be 1, 2, 3, 4, 5, or 6

corresponding to manual LR1 to LR6.

Example

SENS:POW:RANG:LOW 4

See also

SENS:POW:RANG?, SENS:POW:RANG:AUTO,

SENS:POW:RANG:AUTO?, SENS:POW:RANG:HIGH, and

SENS:POW:RANG:LIST?

SENSe:POWer:REFerence

Description This command sets a user-defined reference value (in dBm or

W), for the selected channel when using a PM-1620. If this

command is sent while an acquisition is in progress, the

“Acquisition already running” error message will be raised.

Syntax

SENS:POW:REF

Parameters The parameter is the user-defined reference

value, in the “±99.999” (dBm) or “+999.9999E±99” (W) format.
The parameter represents the unit:
“DBM”sets dBm
“W”sets W

Example

SENS:POW:REF 11.23 dBm