11 Troubleshooting

Problem

Probable Cause

Recommended Action

- - - - - - - displayed as power

value.

Power level is below

the minimum

detectable power.

Select Autorange if a manual

range is selected, or choose a

lower manual range.

+++++++ displayed as

power value.

Power level is above

the maximum

detectable power

(saturation).

Select Autorange if a manual

range is selected, or choose a

higher manual range.
Turn the source off.

! ! ! ! ! ! ! displayed as power

value.

Unstable or

modulated input

signal.

Select a manual range.

Questionable readings

displayed.

Dirty detector or

optical connectors.

Clean the detector and all optical

connections.

Improper wavelength

selected.

Switch to the correct wavelength

on all instruments being used.

Incorrect offset

nulling.

Perform an offset nulling with

protective cap installed.

Unstable optical

source.

Wait for source to stabilize.

Wrong correction

factor.

Reset the correction factor to

+0.000 dB or 1.0000 W/W

High-Speed Power Meter

101

11 Troubleshooting

Problem

Probable Cause

Recommended Action

- - - - - - - displayed as power

value.

Power level is below

the minimum

detectable power.

Select Autorange if a manual

range is selected, or choose a

lower manual range.

+++++++ displayed as

power value.

Power level is above

the maximum

detectable power

(saturation).

Select Autorange if a manual

range is selected, or choose a

higher manual range.
Turn the source off.

! ! ! ! ! ! ! displayed as power

value.

Unstable or

modulated input

signal.

Select a manual range.

Questionable readings

displayed.

Dirty detector or

optical connectors.

Clean the detector and all optical

connections.

Improper wavelength

selected.

Switch to the correct wavelength

on all instruments being used.

Incorrect offset

nulling.

Perform an offset nulling with

protective cap installed.

Unstable optical

source.

Wait for source to stabilize.

Wrong correction

factor.

Reset the correction factor to

+0.000 dB or 1.0000 W/W