Replacing the fuse – EXFO PM-1600 High-Speed Power Meter User Manual
Page 106
Maintenance
96
PM-1600
Replacing the Fuse
Replacing the Fuse
The PM-1600 contains two fuses of type IEC, 250 V, 2 A, fast blow 0.197" x
0.787"/5 mm x 20 mm. The fuse holder is located at the back of the
PM-1600, just beside the power inlet.
To replace the fuse,
1. Unplug the power cord from the PM-1600.
2. Pull the fuse holder out of the PM-1600.
3. Verify and replace the fuses if necessary.
4. Make sure the fuses are firmly in the holder prior to reinsertion.
5. Firmly push the holder into place.
This symbol, found at the back of the PM-1600, indicates that the user
should refer to the instruction manual for fuse replacement and
power rating.
Use only with a 250V fuse
Employer uniquement avec
un fusible de 250V