Controlling the PM-1600 Automatically or Remotely

76

PM-1600

Specific Commands—Quick Reference

READ

[SCAL] POW

DC?

(+999.9999E±99) for

W and W/W units
(±999.9999) for dBm

and dB units

Stores and gets value

ALL

POW

DC?

(+999.9999E±99) for

W and W/W units
(±999.9999) for dBm

and dB units

Returns the power of

both channels in their

respective current unit

SENS

AVER COUN

a

<999>

Sets number for data

averaging

COUN?

(999)

Gets number for data

averaging

[STAT]

a

<0|1>

Sets data averaging

STAT?

(0|1)

Checks if data

averaging is active

CORR COLL

ZERO

a

Performs null

measurement

OFFS

[MAGN]

a

<−5.999 to +6.000>

[DB]
<0.2512 to 3.9811>

[W/W]

Sets correction factor

OFFS

MAGN?

a

(+9.9999E±99) for

W and W/W units
(±9.9999) for dBm

and dB units

Returns correction

factor

FREQ [CONT]

a

[HZ]

Sets low acquisition

rate

CONT?

(999.999)

Gets low acquisition

rate

Command

Parameter/

Response

Description

Controlling the PM-1600 Automatically or Remotely

76

PM-1600

Specific Commands—Quick Reference

READ

[SCAL] POW

DC?

(+999.9999E±99) for

W and W/W units
(±999.9999) for dBm

and dB units

Stores and gets value

ALL

POW

DC?

(+999.9999E±99) for

W and W/W units
(±999.9999) for dBm

and dB units

Returns the power of

both channels in their

respective current unit

SENS

AVER COUN

a

<999>

Sets number for data

averaging

COUN?

(999)

Gets number for data

averaging

[STAT]

a

<0|1>

Sets data averaging

STAT?

(0|1)

Checks if data

averaging is active

CORR COLL

ZERO

a

Performs null

measurement

OFFS

[MAGN]

a

<−5.999 to +6.000>

[DB]
<0.2512 to 3.9811>

[W/W]

Sets correction factor

OFFS

MAGN?

a

(+9.9999E±99) for

W and W/W units
(±9.9999) for dBm

and dB units

Returns correction

factor

FREQ [CONT]

a

[HZ]

Sets low acquisition

rate

CONT?

(999.999)

Gets low acquisition

rate

Command

Parameter/

Response

Description