Managing the wavelength list – EXFO PM-1600 High-Speed Power Meter User Manual
Page 35
Acquiring Power Measurements
High-Speed Power Meter
25
Managing the Wavelength List
Managing the Wavelength List
To view
the list,
1. Press Setup to access the Setup menu.
2. Use the up/down arrow keys to select Wavelength List & Corr. Factor,
then press ENTER. The number of wavelengths in the list is displayed. If
the maximum number of wavelengths (32) in the list is reached, full is
displayed. If the list is empty, empty is displayed. A mark is displayed
next to the active wavelength to indicate that it is on the wavelength
list.
System Setup
Wavelength List & Corr. Factor
Channel 1 Setup
Channel 2 Setup
Exit
Sampling
λ
CF Ch1
CF Ch2
Exit
View
λ List (31 λ)
1537.98 nm
1.0000 W/W
Sort
λ
Add
λ
Clear CF
Clear CF
Del All
Del
λ
+0.000 dB
Active
wavelength
Acquiring Power Measurements
High-Speed Power Meter
25
Managing the Wavelength List
Managing the Wavelength List
To view
the list,
1. Press Setup to access the Setup menu.
2. Use the up/down arrow keys to select Wavelength List & Corr. Factor,
then press ENTER. The number of wavelengths in the list is displayed. If
the maximum number of wavelengths (32) in the list is reached, full is
displayed. If the list is empty, empty is displayed. A mark is displayed
next to the active wavelength to indicate that it is on the wavelength
list.
System Setup
Wavelength List & Corr. Factor
Channel 1 Setup
Channel 2 Setup
Exit
Sampling
λ
CF Ch1
CF Ch2
Exit
View
λ List (31 λ)
1537.98 nm
1.0000 W/W
Sort
λ
Add
λ
Clear CF
Clear CF
Del All
Del
λ
+0.000 dB
Active
wavelength