Remote Control (SCPI) Commands
138
PM-1600
Specific Commands
SENSe:POWer:UNIT
Description This command changes the absolute measurement display
unit: dBm or W (for the selected channel when using a
PM-1620). If this command is sent while an acquisition is in
progress, the “Acquisition already running” error message will
be raised.
Syntax
SENS:POW:UNIT
Parameters The
“DBM” or “0”-set power display to dBm
“W” or “1”-set power display to W (pW, nW, µW, mW...)
Example
SENS:POW:UNIT DBM
See also
SENS:POW:UNIT? and UNIT:POW
SENSe:POWer:UNIT?
Description This query returns the current absolute power measurement
display unit (for the selected channel when using a PM-1620).
Syntax
SENS:POW:UNIT?
Response
“0”-current unit is dBm
“1”-current unit is W (pW, nW, µW, mW...)
Example
SENS:POW:UNIT?
See also
SENS:POW:UNIT and UNIT:POW
Remote Control (SCPI) Commands
138
PM-1600
Specific Commands
SENSe:POWer:UNIT
Description This command changes the absolute measurement display
unit: dBm or W (for the selected channel when using a
PM-1620). If this command is sent while an acquisition is in
progress, the “Acquisition already running” error message will
be raised.
Syntax
SENS:POW:UNIT
Parameters The
“DBM” or “0”-set power display to dBm
“W” or “1”-set power display to W (pW, nW, µW, mW...)
Example
SENS:POW:UNIT DBM
See also
SENS:POW:UNIT? and UNIT:POW
SENSe:POWer:UNIT?
Description This query returns the current absolute power measurement
display unit (for the selected channel when using a PM-1620).
Syntax
SENS:POW:UNIT?
Response
“0”-current unit is dBm
“1”-current unit is W (pW, nW, µW, mW...)
Example
SENS:POW:UNIT?
See also