Section 4 – Controls & Displays

System Status Display

The System Status Display indicates one of the six
system states. If that state is active, the correspond-
ing label is lit in the display. Figure 4-4 shows the
System Status Display with the

ON LINE

state acti-

vated. The six status conditions are described here.

FAULT

This indicator lights when out-of-toler-
ance voltage fluctuations occur that are
related to one of the following condi-
tions.
1. AC Line Voltage
2. Internal DC Power Supply (+5 VDC,

+12 VDC, -12V DC or +24V DC.)

INPUT TEST

This indicator normally flashes red dur-
ing a power-up test. If there is a prob-
lem, a steady red light appears. This
means that either the input AC line volt-
age is out of tolerance, or an internal
fault has occurred in the generator.

OVERLOAD

This red indicator lights when either of
the following conditions occur.
1.

An instantaneous overload caused

by a mismatch between the ultrasonic
signal and the resonant characteristics
of the acoustic stack (transducer,
booster and horn.)

2.

Excessive power beyond the

generator’s rated output is being drawn.

ON LINE

The generator is capable of operation.

OVERTEMP

One of the power modules has over-
heated and the generator has shut down.
This may be caused by excessive dust
in the cooling channel or a cooling fan
failure. The generator will automatically
reset when the module temperature
drops below the trip point which is 75

°

C

(167

°

F).

OFF LINE

The generator is in a standby mode. The
ultrasound output cannot be activated.

ON LINE

OVERTEMP

OFF LINE

FAULT

INPUT TEST

OVERLOAD

Figure 4–4 System Status Display