The assembly and disassembly procedures for a
hand probe shown in Figure 10–6. The same pro-
cedure for a probe stack with booster is shown in
Figure 10–7. It makes no difference whether the
horn is attached to the booster first, or the booster
is attached to the probe first.

PROBE
HOUSING

BOOSTER

HORN

Ultrasound
Coaxial
Connector

Mounting Stud

Input End

Spanner

Wrench

Hole

Spanner

Wrench

Hole

(Typical)

Output End

(Threads Deeper
Than Input End)

Booster Input End

(Threads Shallower
Than Output End)

(Cap Screws Face
Toward Input)

Booster
Mounting
Ring

(Do not clamp)

Booster
Mounting
Ring

(Do not clamp)

Input End

Output End

Mounting Stud

(Insert Into Horn
First and Tighten
to Torque Specs)

(Insert Into Booster Input
End First and Tighten
to Torque Specs)

PROBE
STACK
ASSEMBLY

A

SSEMBLY

D

ISASSEMBLY

Figure 10–7

Probe Stack with Booster Assembly and Disassembly

ƽ

CAUTION

Never hold a probe by
the housing or a booster
by the mounting rings
when tightening or loos-
ening an adjoining com-
ponent. The probe
housing and booster
rings have anti–rotation
devices to keep the
transducer and booster
aligned and could shear
under excessive torque.