Section 12 – Troubleshooting
INPUT TEST
indicator
lights up
?
INPUT TEST
flashes for
5–10 sec
?
1. Both the front and rear AC breaker switches on
the DPC must be turned on (Figure 3–9 and 4–2).
2. Make sure the AC outlet has power.
YES
NO
YES
NO
Verify that the AC voltage is within specifications.
INPUT TEST
indicator
goes out
?
YES
NO
Verify that the AC voltage is within specifications.
START
Green
AC power
light on
?
YES
NO
1. Defective power indicator lamp.
2. Internal fault – contact Dukane Service.
FAULT
indicator
off
?
YES
NO
Verify that the AC voltage is within specifications.
A
(Continued on next page)
(Plug in AC Powercord into outlet)
Figure 12–3
Troubleshooting Flowchart – Part 1