8^2,16 Zero Offset Adjustment

Zero offset can be applied lo make the flowmeter ouipiits comparable lo process values

measured by other instruments. If the ¿его adjustment described in 6.2 requiring a zero

flow rate condition can he performed, diis zero offset adjustment is not needed. When the

zero adjustment is completed, zero offset will be automaiically cleared to zero.

Zero offset can be set in the range dciicribed below:

Zero oiiset: ±0,125 m/s (±1.25

7c

of 10 m/s—maximum range) maximum

IVocecd as follows to check or change the zero offset value.

■ I'o check the zero oHiset value:

Key operation

Display example

Description

SKI

U1: MANUAL ZERO

tOG2. 5

%

Press [SET] first to start the items selection sequence

and select Ml: MANUAL

ZERO

from among the

configuration items using [ ► ] and [

a

] keys. Then

press fSET] ufiain to display the current zero offset value

SET

Ml: MANUAL ZERO

Pressing [SETj, the system returns to the items selection

sequence.

I To change the zero offset value:

Calculate the zero offset value with the following equation:

Zero offset value (%) = {{actual flow rate) - (LF404 measured value)}

The zero offset value should be c:dculated in percent value for Range 1. See the following

example.

(Example)

Measured condition

Flow rate

%

in measurinf! span

Actual flow' rate obtained

from other insirumem,

10,0

m-V'min

50 %

LP4U4 measured value

10.5 m Vmin

52.5

%

Zero offset

-2.5 %

If zero offset is set lo -2.5 %, the LF404 converter will output 50,0 %
llnvr' rate instead of.52,5%,