15. Specifications

The flowmeter specification:s and Ihc type specification code used when ordering the

flowmeter are described in this chapter.

15*1 Flowmeter 5 pecifi cations

■ Overall Specifications

Measurement range m terms of flow velocity:

0-(J3 m/s to 0-10 in's (0-0.1 nvs to 0-0.3 nvs range is available optionally)

System accuracy:

See the following table.

Table 15.1 System accuracy

Flow rate as a
percentage

of

range

Accuracy

O.l-LO m/s OJ -

1*0

m/s

1

*

0-10

m/s

0

to

20

%

±0.1% FS

20

to

100

%

±0.5% of rate

0 to 50%

iO.25% of FS

50 to 100%

±0.5% of rate

Note: The accuracy above is measured under standard operating conditions at

Toshiba’s calibration facility.

Fluid conductivity:

5 ,«S/cm minimum

Fluid tempera! ure:

-10 to -i-120 " C

Ambient temperature: -10 to H-fiO

° C

Dimensions and Mass; See Chapter 16, “Outline Dimensions,’'

-

1 43 -