10.8 Remote Selection of Fixed Value Output

A

List

rasped fied current output anti pulse output can be selected with a Dl signaL

Proceed as follows to use this function:

Fixed-value setting

Set the fixed-value for current output and for pulse output. See S,2.L%

'Tixed‘Viiluc Output.'" Set the fixed-value output cnabk/disable status to ‘"OFF7^

If the pulsse output is not used, fixed-value setting for puUe outpul is not needed.

Dl function setting

Set Dl to use as a fixed-value output control signal. See 8.2.9, ‘‘Digital I/O ”

Control signal input conditions:

Control signal Input level

4 -20 niA and pulse output

L level

Outputs the measured value.

H level

Outputs the fixed-value.