8 remote selection of fixed value output, Toshiba – Toshiba LF494 User Manual
Page 127
Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".
TOSHIBA
6 F 8 A 0 7 7 , 4
10.8 Remote Selection of Fixed Value Output
A
List
rasped fied current output anti pulse output can be selected with a Dl signaL
Proceed as follows to use this function:
Fixed-value setting
Set the fixed-value for current output and for pulse output. See S,2.L%
'Tixed‘Viiluc Output.'" Set the fixed-value output cnabk/disable status to ‘"OFF7^
If the pulsse output is not used, fixed-value setting for puUe outpul is not needed.
Dl function setting
Set Dl to use as a fixed-value output control signal. See 8.2.9, ‘‘Digital I/O ”
Control signal input conditions:
Control signal Input level
4 -20 niA and pulse output
L level
Outputs the measured value.
H level
Outputs the fixed-value.
- 1 2 6 -