(6)Groiindirif<

The grounding terminal of the LF494/LK404 Q

o

a

meter should be grounded with lOUohm

or less ground resistance. Use a heavy copper braid or wire (cross-sectional area

5 , 5

nun"

minimum) to ground the lerminal and make it as short as possible. The terminal is M4 siiic
and an M4-siiie crimped ring fug should be used to connect the wire to the terminal. Avoid a

common ground where earth currcni may fliw. An independent ground is preferable, See

Figure 4.1(1.

To prevent a iwo-poini grounding, ground the shielded cable on the receiving instrument

side,

Figure 4.10 Grounding Procedure