beautypg.com

Toshiba – Toshiba LF494 User Manual

Page 134

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

background image

TOSHIBA

6F.8A0 7 74

Converter hardware failure

The system checks the internal circuitry at the time of power-up for all error items end

checks continuously for the specified items as described below, if an error is detected, one

of the messages shown in the table below will be displayed.

If multipJe errors occur, their messages will be displayed cyclically* The diagnostics items

concerning the excitation cable and excitation circuit are detected using the АПС circuit.

Thus, if the ADC fails, No. 4 (excitation cable) and No. 5 (excitation circuit) errors can

not be Jciccted correoily. Further, this entire checking system is based on the CPLi in the

llownieter. Therefore, if the CPU fails, no accurate diagnostics or error message display

can be obtained*

No.

Error message

Description

Corrective action

1

t ROM ERROR *

ROM error

Internal components or

printcd-circuit board must be
repaired or replaced.
Contact your nearest Toshiba
representative*

2

* RAM ERROR *

RAM error

3

PARAMETER FAIL

System parameter error

4

El CURR* OPEN

Excitation cables are not
connected,

Connect the excitation cables
corre clly.

5

EL GURU ERROR

An error occurred in the
excitation circuit.

Internal components or
prinled-circuit board must be
repaired or replaced.
Contact your nearest Toshiba
representative.

6

ADC. ERROR

ADC error

7

INVALID TOTAL

Totalizer tala was destroyed
due to external noise. (No
message appears if totalization
is not used.)

The error message disappears

if you press the reset key.

NOTES

T

Errors No. 1 to No. 3 can be detected only at the tinte of power-up. llie flowmeter

does not start measurement if any one of these errors is delected. If these errors

occur after power-up, the flowmeter cannot detect these errors, and thus Eiiay

indicate and output incorrect data.

2.

Errors No, 4 to No,

6

may not be delected even if the errors result in incorrect

flowmeter accuracy> because of characteristic differences in components used to

detect these errors.

3.

CPU error cannot be deice led. If the CPU siop.s, the watchdog timer resets the
internal circuits and the flowmeter starts again from the initial power-up condition.

Depending on CPU condition, the flowmeter may not indicate and output correct

data.

-

1 3 3

-

This manual is related to the following products: