Setting Alarm Limits

Propaq Encore Reference Guide

99

Welch Allyn

Turning On and Off All Limits

If you want to turn all limits on or off without changing their values, press

ALL ALRM

in the

Alarms Menu (a confirmation window appears requiring you to confirm your selection). The
Alarms Status Window displays bells when alarms are on.

Changing Individual Limits

Setting each limit requires selecting the desired vital sign parameter window, selecting the
limit to adjust, and changing the limit.

1.

From the Main Menu, press

SETUP

,

ALARMS

.

2.

Press

LIMITS

to display the alarm limits window and the Limits Menu.

3.

Press

NEXT PAGE

to change to the desired alarm limit window.

4.

Press the

NEXT

button to move the cursor.

5.

Press

UP

or

DOWN

, or

ON/OFF

(if

ON/OFF

is available) to set the limit to the desired limit

value.

6.

When the limit is set, select the next limit with the

NEXT

button. Or, to select another

vital sign, press

NEXT PAGE

.

Repeat steps 4 through 6 to adjust the desired limits. Continue the process until you've set all
the limits you want.

Warning If a patient’s vital sign value falls outside of the upper or lower alarm
range limit,

STAT SET

turns off the alarm and the alarm limit except for the

following:

1. The lower alarm limits for SpO

2

and ETCO

2

are not turned off by

STAT SET

.

2. If HR/PR ALARM LIMITS in the Settings window is set to CANNOT TURN

OFF,

STAT SET

affects HR/PR alarm limits as follows: for an overrange HR/

PR patient value (displayed as +++), the upper alarm limit is set to the
maximum alarm limit, and the lower alarm limit is unchanged; for an
underrange HR/PR patient value (displayed as - - -), the lower alarm limit
is set to the minimum alarm limit and the upper alarm limit is unchanged;
for an indeterminate HR/PR patient value (displayed as ???), the alarm
limits are unchanged.