Battery Care

Propaq Encore Reference Guide

123

Welch Allyn

Battery Care

Recharging Time

The battery charges to full capacity within eight hours (if the monitor remains off).

Monitor Functions Resumed

In most instances, most monitor functions are usable immediately after plugging in the ac
power adapter and cycling the power switch. More charging time may be required before the
NIBP, CO

2

, and printer can be operated.

Operating Times Using Battery Power

The amount of time you can operate the Propaq Encore on each battery charge depends upon
many variables including active options, frequency of NIBP measurements, frequency and
length of print strips, ambient temperature, battery age and condition, and what information
is displayed.

For monitors without the Expansion Module or SpO

2

option, typical monitor operating time

is about 2 hours at 25° C for a new, fully-charged battery. This is when all patient channels are
active and measurements are taken every 15 minutes.

For monitors without the Expansion Module but with the SpO

2

option, operating time is

about 5 hours.

For monitors with the Expansion Module and printer, SpO

2

and CO

2

options, typical

operating time is about 3 hours under the above conditions when print strips are generated
every 15 minutes.

Caution Leaving the monitor's lead-acid batteries in a completely discharged
state may result in permanent battery damage. The batteries should be kept
fully charged.