Life-threatening alarms, Temporary alarms, Life-threatening alarms temporary alarms – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual
Page 94: Art mmhg

Types of Propaq Alarms
Welch Allyn
94
Propaq Encore Reference Guide
The alarm tones are summarized in the table:
Life-Threatening Alarms
Life-threatening alarms are automatically turned on when the channel is activated, and
cannot be turned off by the user. A life-threatening alarm is the highest priority patient
alarm. You are notified of any life-threatening alarm as soon as it is detected. An apnea alarm
is a life-threatening alarm.
Temporary Alarms
A temporary alarm occurs when a patient's condition changes, causes an alarm to occur,
changes again, and no longer violates a limit.
After a temporary alarm, review each vital sign’s alarm limit window until you find the limit
with an asterisk (*) next to it. This limit caused the alarm. The asterisk(s) are cleared when
you leave that limits window.
1.
Press
ALARMS
and then
LIMITS
to display the first limits window.
2.
Press
NEXT PAGE
(in the event of multiple alarm violations) until all asterisks are
located.
Alarm Condition
Tone
Patient alarm
Continuous ON
Apnea alarm
ON for 1 second, OFF for 1 second
Equipment alert
ON for 1 second, OFF for 4 seconds
mmHg
NEXT
NEXT PAGE
UP
MCO
2
BR
SpO
2
PA
ALARM LIMITS
35/ 18
( 85)
35 12
92
DOWN
S :
141
112
D : 95 * 62
(M) : 102 68
UPPER LOWER
ART
mmHg
ON/OFF
Alarm Limits
Menu
Alarm Limits
Status
Window
Asterisk indicates
a temporary alarm
limit violation