NIBP

Welch Allyn

60

Propaq Encore Reference Guide

3.

Screw the hose connector onto the NIBP connector on the monitor's left side.

4.

If motion artifact such as shivering, coughing, or vehicle motion interferes with NIBP
readings, do the following:

Position the patient’s limb away from the body so the applied cuff is not in contact
with the patient’s body or any other object such as a bed rail. Try to keep the cuff at
the same level as the heart.

Connect the ECG leads to the patient and perform ECG monitoring during NIBP.

Set Up the NIBP Channel

Press the

NIBP

button to display the status window and menu.

A cuff that is not properly connected to the patient may result in a false
reading if the patient and cuff are moved by motion artifact or clinical
personnel during the NIBP measurement. Always verify the cuff is properly
placed on the patient.

START/STOP

Starts and stops NIBP measurements. Any time the Propaq Encore is taking a
noninvasive pressure measurement, the

START

button changes to

STOP

so you

can stop the measurement in progress. This button initiates the same action as
the

START/STOP

button at the left side of the screen. Pressing

STOP

will

automatically vent the cuff.

STOP

SMARTCUF

INTERVAL TURBOCUF

AUTO/MAN

ART

HR

T

1

NIBP

18:46:00

18:45:28

III

1

mV

cm

MON

NIBP ADULT

MODE : TURBOCUF
TIME : 4:21
SMARTCUF : ON

102.3

°

F

S

122

( 87)

85

109

72

122

58

80

85

50 100 150 200

85

122

58

NIBP

99

SpO

2

mmHg

mmHg

122

58

85

This symbol indicates the NIBP reading was taken

in the presence of high motion artifact while

monitoring ECG. Artifact can affect accuracy.

To help reduce artifact, see page 60.

NIBP

122/

58

( 85)

18:45

Manometer Bar.

Systolic, diastolic,

and mean values

are indicated above

the manometer as

small triangles.

Time of last measurement

Last measurement

Measure-

ment mode

Alarm Limits Bell

NIBP values and units

of measure rotate in

and out of this window

for up to one hour

(International English

only)

NIBP values are

displayed in one of

these windows if

vacant. After 16

minutes, the display

changes to that

shown below for 44

minutes.

Time

remaining in

Turbocuf

Mode or Auto

Mode interval