Noise on the signal, Using the filter to better display a waveform, Ecg messages – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 48

background image

ECG/RESP

Welch Allyn

48

Propaq Encore Reference Guide

Noise on the signal

Noise on the ECG signal may be detected as pacer signals, causing the pacer indicator to
appear on the display. If you don't need to indicate pacemaker signals, you may want to turn
off the pacemaker indicator for a better display of the ECG waveform.

Using the Filter to Better Display a Waveform

The Propaq Encore includes a filter that reduces noise from the ac power signal and produces
a much clearer ECG waveform. To set the filter, press

SETUP, MORE, MORE, SERVICE, YES

(to access

the Service Menu),

MORE, MORE, SETTINGS

to display the Settings Menu. Select the FILTER

function with the

NEXT

button and press the

CHANGE

button to change settings.

Set the filter to the ac mains frequency of your hospital (either 60 or 50 Hz).

ECG Messages

If a lead fault occurs, an ECG equipment alert typically shows which lead failed. Otherwise,
the Propaq Encore displays MULTIPLE in place of the failed leads.

ECG LEAD CHANGED

. The Propaq Encore has automatically changed an ECG lead

due to a lead wire or electrode problem.

LEAD FAIL: REPLACE ELECTRODES.

There is excessive offset on the ECG channel.

Warning The presence of much pacer-like noise can cause the displayed
heart rate to be erratic even though the ECG trace may look clean with the
pacer indicator off. Fix the noise problem by using fresh ECG electrodes and
an ECG cable whose lead wires make good connections.