Capnography (CO2)

Propaq Encore Reference Guide

79

Welch Allyn

Sidestream CO

2

Menu and Status Window

When Sidestream CO

2

is active, press

MORE

from the first CO

2

menu to access the SCO

2

menu and the SCO

2

status window:

SWEEP SPEED

The CO

2

and RESP display sweep speed (set by the

mm/s

button)

BAROMETER

The ambient barometric pressure

To change

CO

2

units, use the Settings Menu under the Service Menus.

Changing these units will clear

CO

2

trends.

GAS COMP

Selects the measurement compensation for CO

2

measurements

RESPONSE

Sets the response time for CO

2

measurement (NORMAL, FAST, OR SLOW)

SOURCE

Changes between Mainstream CO

2

and Sidestream CO

2

monitoring (if both

options are installed), or disables CO

2

monitoring

Setting CO2 SOURCE to OFF allows you to disable CO

2

monitoring without

removing the sensor. When the CO2 SOURCE is set to OFF, the Propaq
displays OFF for CO

2

numerics.

SWEEP SPEED

The CO

2

and RESP display sweep speed (set by the

mm/s

button)

BAROMETER

The ambient barometric pressure

FLOWRATE

Sets the sampling flow rate to either 90 or 175 ml/min. You can change the
flow rate while Sidestream CO

2

is active

To change

CO

2

units, use the Settings Menu under the Service Menus.

Changing these units will clear

CO

2

trends.

mmHg

GAS COMP

PREV MENU

RESPONSE

SCO

2

BR

SpO

2

PA

CO

2

GAS COMPENSATION: OFF
RESPONSE : NORMAL
CO2 SOURCE : SIDESTREAM
SWEEP SPEED : 6.25 mm/s
BAROMETER : 762.0 mmHg

35/ 18

( 85)

35 12

92

SOURCE

FLOWRATE : 175

ml/min

FLOWRATE