Impedance pneumography selections – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 46

background image

ECG/RESP

Welch Allyn

46

Propaq Encore Reference Guide

When you press

MORE

in the first ECG/RESP Menu, a status window appears showing you

the current ECG/RESP settings and additional selections.

The Size function “increases” or “decreases” the ECG or RESP waveform size. Each time you
press a

SIZE

button, the waveform approximately doubles in height. When you reach the

largest waveform size, the next press displays the smallest size.

Impedance Pneumography Selections

If your Propaq Encore includes the Impedance Pneumography Option, the following
selections are also available.

HR/PR TONE

Sets the heart tone loudness to LOW, MEDIUM, HIGH, or OFF.
If SpO2 is monitored, the pitch of the tone varies with the SpO2 value.

PACER DISPLAY

Turns on and off the pacer indicator in the ECG waveform.

ECG

BANDWIDTH

This selection allows you to determine the bandwidth for the data sent
to the display and the printer. If the selection is Extended, the
bandwidth is 0.05-40 Hz. If the selection is Monitor, the bandwidth is
0.5-40 Hz.

The QRS detector sensitivity threshold is not affected by changing the ECG
display size. Likewise, the RESP breath detector threshold is not affected by
changing the RESP display size.

RESP LEAD

Selects the RESP lead. Choices are RA-LA and RA-LL, and choice is
independent of ECG lead selection. Experiment with placement for best
signals.

RESP

Turns impedance pneumography (RESP channel) on or off.

NIBP

mmHg

NEXT

PREV MENU

CHANGE

MCO

2

BR

SpO

2

PA

HR

T

1

NIBP

18:45:28

III

1

mV

cm

MON

ECG/RESP
HR/PR TONE :

LOW

PACER DISPLAY: ON
ECG BANDWIDTH: MONITOR (0.5-40Hz)
RESP LEAD : Ld1 (RA-LA)
RESP : ON

102.3

°

F

35/ 18

35 12

92

80

( 85)

122

58

Selected ECG

lead

Sensitivity

factor

Bandwidth

Selection

Available with the

Impedance

Pneumography

Option