beautypg.com

Power, Monitor power specifications – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 155

background image

Power

Propaq Encore Reference Guide

155

Welch Allyn

Power

Monitor Power Specifications

Electromagnetic Compatibility (EMC) Per IEC 601-1-2, which is a collateral standard of IEC 601-1, for

electromagnetic compatibility.

Paper

Short-term Storage Environment
(up to 7 days)

-20 to 40°C; 5% to 80% noncondensing

Long-term Storage Environment
(up to 5 years)

25°C (optimal), 65% noncondensing

Characteristic

Specification

Mode of Operation

Continuous

Battery Pack Type

Sealed, gel-type lead acid

Battery Pack Capacity

Monitor only: 8 volts, 3 Ampere-Hours; Monitor with Expansion
Modules: 8 volts, 6 Ampere-Hours

Battery Recharger Circuitry

Internal, powered by external power adapter

DC Input Power Required

12 to 28 Volts, 25 Watts

Input Fuse Rating

3A/250V, Type 2AG (0.57x 0.177 in)

Operating Times on Battery

Monitor only

Monitor and SpO2 (Baqpaq)

Monitor with Expansion Module with
printer, SpO2 and CO2 Options

2

.5 hours

4.5 hours

3 hours

Battery Recharge Time with
instrument on

Range of 8 hours to 12 hours typical, depending upon product
configuration

Battery Recharge Time with
instrument off

Range of 6 hours to 8 hours depending upon product configuration

Recharge time until monitor is
usable, starting with discharged but
non-faulty battery

≤ 2 minutes typically (longer time required before NIBP, printer, and
CO

2

are available)

Characteristic

Specification