Monitor Setup
Welch Allyn
30
Propaq Encore Reference Guide
HR/PR TONE
Sets the Heart Tone volume to HIGH, MEDIUM, LOW, or OFF.
PATIENT MODE
Pressing
CHANGE
in this selection displays the following Patient Mode
window:
If you press
NEONATAL
,
PED
, or
ADULT
, a confirmation window
appears, requiring you to confirm your selection:
Whenever you change the patient mode, the alarm limit
settings are automatically changed to the defaults for that
mode. If Custom settings have been set for that mode, the
defaults are the Custom mode settings. If no Custom settings
have been set, the defaults are the Factory Mode settings.
See page 33 for more information about patient modes.
If you change the patient mode, the CO
2
alarm limits in the
new mode might vary slightly from the originally-programmed
CO
2
limits for the new mode. Check the CO
2
alarm limits.
If you press
SETUP
in the previous Patient Mode window, the
Mode Setup window appears. This allows you to set custom
patient modes and powerup defaults as described on
page 33.
SETUP
BATTERY: 9.2 VOLTS
mmHg
NEONATAL
PREV MENU
ADULT
PED
MCO
2
BR
SpO
2
PA
PATIENT MODE
SELECT PATIENT MODE BASED ON AGE:
NEO : < 44 WEEKS GEST. AGE
PED : > 44 WEEKS GEST. AGE,
< 9 YEARS
ADULT: > 9 YEARS
35/ 18
( 85)
35 12
92
mmHg
YES
NO
MCO
2
BR
SpO
2
PA
PATIENT MODE
CHANGING SETTINGS AND MODE
35/ 18
( 85)
35 12
92
ARE YOU SURE?
TO:
ADULT ALARM LIMITS WILL BE SET
CUSTOM: ADULT