Monitor Setup

Welch Allyn

30

Propaq Encore Reference Guide

HR/PR TONE

Sets the Heart Tone volume to HIGH, MEDIUM, LOW, or OFF.

PATIENT MODE

Pressing

CHANGE

in this selection displays the following Patient Mode

window:

If you press

NEONATAL

,

PED

, or

ADULT

, a confirmation window

appears, requiring you to confirm your selection:

Whenever you change the patient mode, the alarm limit
settings are automatically changed to the defaults for that
mode. If Custom settings have been set for that mode, the
defaults are the Custom mode settings. If no Custom settings
have been set, the defaults are the Factory Mode settings.
See page 33 for more information about patient modes.

If you change the patient mode, the CO

2

alarm limits in the

new mode might vary slightly from the originally-programmed
CO

2

limits for the new mode. Check the CO

2

alarm limits.

If you press

SETUP

in the previous Patient Mode window, the

Mode Setup window appears. This allows you to set custom
patient modes and powerup defaults as described on
page 33.

SETUP

BATTERY: 9.2 VOLTS

mmHg

NEONATAL

PREV MENU

ADULT

PED

MCO

2

BR

SpO

2

PA

PATIENT MODE
SELECT PATIENT MODE BASED ON AGE:

NEO : < 44 WEEKS GEST. AGE
PED : > 44 WEEKS GEST. AGE,
< 9 YEARS
ADULT: > 9 YEARS

35/ 18

( 85)

35 12

92

mmHg

YES

NO

MCO

2

BR

SpO

2

PA

PATIENT MODE

CHANGING SETTINGS AND MODE

35/ 18

( 85)

35 12

92

ARE YOU SURE?

TO:

ADULT ALARM LIMITS WILL BE SET

CUSTOM: ADULT