Alarms
Alarms Specifications
Characteristic
Specification
Indicators
Flashing red ALARM light indicates patient alarm; continuously on
ALARM light indicates patient alarms are suspended. Continuously on
ALARM(S) OFF light indicates one or more alarm limits have been
disabled. A flashing ALARM(S) OFF light indicates an equipment alert.
Tone Frequency
900 Hertz
Tone is steady for a patient alarm and sounds for 1 second every 4
seconds for an equipment alert.
Selectable Tone Volume
Low, Medium, High
Limits
Settable on all parameters
Control
Automatic preset or manual settings
Alarm on Tachycardias
Most tachycardias will alarm in less than 8 seconds. These include
AAMI 3.1.2.1 part f. waveforms. Certain multifocal tachycardias may
initially alarm as “low rate.”
Apnea delay setting
Adult/Ped = 6, 10, 15, 20, 25, 30 seconds
Neonate = 6, 10, 15, 20 seconds
Alarm Holdoff Time Period
a
a.Alarm holdoff time period is reset if the vital sign returns to acceptable limits before an alarm occurs.
HR/PR = 3 seconds (except NIBP PR)
SpO
2
= 10 seconds
RR/BR = 5 seconds
Audio Alarm Holdoff with Acuity
When a Propaq Encore in Adult or Pediatric Mode is connected to an
Acuity System, the audio alarms at the bedside Propaq can be delayed
up to 4 minutes and 15 seconds. The delay time is selected in Acuity
software at the time of Acuity installation. Visual alarm indications are
not delayed.