beautypg.com

Power adapters, Power adapter general specifications – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 156

background image

Power Adapters

Welch Allyn

156

Propaq Encore Reference Guide

Power Adapters

Power Adapter General Specifications

Characteristic

Specification

Protection Classifications, all Adapters

a

a.Per EN 60601-1 unless otherwise stated.

Type of Protection Against Electric
Shock

Class I, (Protectively Earthed)

Degree of Protection Against Harmful
Ingress of Water

For ordinary, indoor locations only.

Method of Disinfection

Not suitable for autoclaving

b

b.See Chapter 8 for cleaning instructions.

Flammable Anesthetics

Not suitable for use with flammable anesthetics

Environmental Specifications, All Adapters

Operating Temperature

0° to 50° C

Shipping and Storage Temperature

-20° to 60° C

Operating Altitude

-2,000 to 15,000 feet (-610 to 4,572 m)

Shipping and Storage Altitude

-2,000 to 40,000 feet (-610 to 12,192 m)

Operating Relative Humidity

15% to 95%, noncondensing

Shipping, Storage Relative Humidity 15% to 95%, noncondensing

Shock

50 g

Vibration

Random Vibration, 0.02g

2

/Hz from 10 to 300 Hz, ramping down to

0.002g

2

/Hz at 500 Hz. Operating 1 hour per axis,

3 hours per test.

Water Resistance

For ordinary, indoor locations only.