beautypg.com

Nibp, Nibp specifications – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 142

background image

NIBP

Welch Allyn

142

Propaq Encore Reference Guide

NIBP

NIBP Specifications

Characteristic

Specification

Method

Oscillometric

Control

Automatic and manual measurement control

Auto Intervals

1, 2, 3, 5, 10, 15, 30, and 60 minutes

Turbocuf

Maximum measurements allowable in a 5-minute period

Displayed Pressures

Systolic, Diastolic, and Mean plus on-screen manometer

Systolic Range

Adult: 30 to 260 mmHg
Ped: 30 to 160 mmHg
Neonate: 25 to 120 mmHg

Diastolic Range

Adult: 20 to 235 mmHg
Ped: 15 to 130 mmHg
Neonate: 10 to 105 mmHg

Mean Range

Adult: 20 to 255 mmHg
Ped: 15 to 140 mmHg
Neonate: 10 to 110 mmHg

Static Manometer Accuracy

±3 mmHg

Minimum Inflation Pressure

Adult: 100 mmHg
Ped: 80 mmHg
Neonate: 50 mmHg

Maximum Allowable Pressure

Adult: 270 mmHg
Ped: 170 mmHg
Neonate: 132 mmHg

Default Inflation Pressure

Adult: 160 mmHg
Ped: 120 mmHg
Neonate: 90 mmHg

Normal Overpressure Limit (results in
up to 2 retries)

Adult: 280 mmHg
Ped: 200 mmHg
Neonate: 141 mmHg

Single Fault Overpressure Limit

Adult: 308 mmHg
Ped: 220 mmHg
Neonate: 154 mmHg

Leak Rate

After a 1 minute settling period, leak rate is

≤4mm/Hg over a 3-minute

period at 270 mm/Hg.

Pulse Rate Range

30 to 220 beats per minute

Maximum Determination Time (with
retries)

Adult: 4.5 minutes
Ped: 4 minutes
Neonate: 3 minutes

Maximum Determination Time (no
retries)

Adult: 3 minutes
Ped: 2 minutes
Neonate: 1.5 minutes

Typical Determination Time without
Artifact

30 to 45 seconds