beautypg.com

Printer fault messages – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 36

background image

Printer Functions

Welch Allyn

36

Propaq Encore Reference Guide

PRINTER FAULT Messages

These PRINTER FAULT messages can appear in an equipment alert window.

The front panel of the printer lets you control the basic printer functions.

If you press

FREEZE

prior to pressing

SNAPSHOT

, the printer prints the 8 seconds of patient

information obtained prior to when you pressed

FREEZE

.

LOW BATTERY,
PRINTER DISABLED

This message appears when the Propaq’s battery voltage is
less than 7.6 volts. To continue operation, plug the ac power
adapter into the Propaq.

CHECK DOOR

The door on the bottom of the printer is open. Close door to
remove this message.

PAPER OUT

T

o add printer paper, see page 128.

OVERHEATING

The printer is overheating. Service may be required.

START
STOP

SNAP
SHOT

PRINT
TRENDS

Manually starts and stops a printout of
patient information as it is monitored
(continuous or real time).

Hold down top two keys simultaneously to
generate a paper feed.

Hold down top
and bottom keys
simultaneously to
generate a
test strip.

Prints all trends that are enabled in the
Printer Trend Select Window.

Prints the last 8 seconds of data for
nonrespiration waveforms and 32
seconds of compressed waveform history
for respiration waveforms.