Capnography (CO2)

Propaq Encore Reference Guide

75

Welch Allyn

Capnography (CO

2

)

Intended Use

The Propaq’s Capnography (CO

2

) option is intended to noninvasively measure the following

vital signs or events: End-tidal CO

2

(ETCO

2

), Inspired CO

2

(INCO

2

), Breath Rate, and Apnea.

The CO

2

option is available as Mainstream CO

2

, Sidestream CO

2

, or Dualstream

CO2

.

Although Dualstream CO

2

provides both Mainstream CO

2

and Sidestream CO

2

monitoring,

only one method can be used at a time. The v option is required for any Propaq equipped
with the CO

2

option.

CO

2

reading accuracy is affected by the presence of interfering gases and vapors. If the CO

2

option is used on patients who are being administered oxygen (O

2

) or nitrous oxide (N

2

O),

be sure to set the appropriate compensation setting using the

GAS COMP

button.

Warning Before you use a Propaq on a new patient, always turn it off for a
few seconds, then turn it on again. This clears the prior patient’s trend values,
alarm limit settings, and NIBP cuff inflation target.

Avoid exposing a Propaq with the

CO

2

option to non-patient sources of

CO

2

such as vehicle engine exhaust or smoke. When such exposure is possible,
avoid opening the printer door. Exposure to these CO

2

sources can

temporarily trap v within the monitor or Mainstream v sensor housing, even
when monitor power is off. This can temporarily cause an erroneous elevated

CO

2

measurement baseline until the trapped

CO

2

leaks out and the baseline

returns to zero (which can require as long as 3-24 hours).

Caution The Mainstream v and Sidestream

CO

2

options are not

recommended for use during magnetic resonance imaging (MRI) procedures.
The magnetic fields involved will permanently damage the CO

2

sensor.

CO

2

monitoring outside the specified operating temperature range can cause

inaccurate

CO

2

readings. The operating temperature range for the

CO

2

option

is different than the range of 0° to 40° C for other Propaq Encore functions:

Mainstream

CO

2

operating temperature: 10° to 40° C

Sidestream

CO

2

operating temperature: 5° to 40° C