Monitor (Environmental)

Welch Allyn

152

Propaq Encore Reference Guide

Monitor (Environmental)

Monitor Environmental Specifications

Characteristic

Specification

Operating Temperature

0° to 40° C

Shipping and Storage Temperature

-20° to 60° C

Operating Altitude

-2,000 to 15,000 ft (-610 to 4,572 m)

Shipping and Storage Altitude

-2,000 to 40,000 ft (-610 to 12,192 m)

Operating Relative Humidity

15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural

Shipping and Storage Relative
Humidity

15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural

Shock

50 g

Vibration, Random

0.02g

2

/Hz from 10 to 500 Hz, ramping down to 0.002g

2

/Hz at 2000 Hz.

Operating 1 hour per axis, 3 hours per test. Designed to meet RTCA
DO-160C, Category C.

Vibration, Sinusoidal

0.10 peak to peak inches 5 to 17 Hz, sloping to 0.01 peak to peak
inches at 55 Hz, then sloping to 0.0001 peak to peak inches at 2000
Hz. Operating 1 hour per axis, 3 hours per test. Designed to meet
RTCA DO-160C, Category N.

Electromagnetic Compatibility (EMC) Per IEC 601-1-2, which is a collateral standard of IEC 601-1, for

electromagnetic compatibility. Designed to meet RTCA DO-160C,
Section 21, Category B. Propaq 200 Series SN EA000225 and higher:
MIL-STD 461D:CE102, CS101, CS114, CS115, CS116, RE102,
RS103 (United States Air Force).

Caution The monitor may not meet performance specifications if it is not
used or stored within these environmental specifications.